OLYMPUS CF-H190L colonoscope SN: 240****.

Test results:

1. Image ok;

2. Light bundle still ok;

3. The front end of the insert tube is broken, and the serpentine;   the position of stress boot is leaking;

4. Biopsy tube wear;

5. Left Angle failure, bending section, wire and helix tube wear;

6. Soft and hard adjustment wire wear, relatively tight;

7. Rubber leakage.

Maintenance:

Replace the insertion tube, biopsy tube, bending section, wire, helix tube, soft and hard adjusting wire and rubber. (No warranty for CCD)