Medtronic 1898200T M4 Straightshot Microdebrider and Handpiece for repair
Medtronic 1898200 M4 handpiece, SN:5***.
Fault: 1) The bearing is corroded and the seal is broken.
2) Motor stuck, inside is full of disinfection residue, motor corrosion.
3) The rotation direction of the knife head is stuck.
4) Handle paint off, internal corrosion.
Solution: Exchange the whole handpiece except cable.
Repair cost: $****.
Warranty: 90 days.



Medtronic 1898200T M4 Straightshot Microdebrider and Handpiece for repair
Medtronic 1898200 M4 handpiece, SN:HP-2****.
Fault: 1) The bearing is corroded and the seal is broken.
2) Motor stuck, inside is full of disinfection residue, motor corrosion.
3) The rotation direction of the knife head is stuck.
4) Handle paint off, internal corrosion.
Solution: Exchange the whole handpiece except cable.
Repair cost: $****.
Warranty: 90 days.



We provide trading and service of Ultrasound unit/Transducer, Dynamic system device/parts, Rigid endoscope parts and Flexible endoscope parts, we also provide trading and service of Surgical Instrument like ureteral renal forceps, electrocoagulation hook, mirror sheath, gravel needle, pneumoperitoneum needle, puncture needle, rongeur, separation forceps, separation scissors, biopsy forceps, operating hand, spreader, biopsy soft forceps, titanium clamp , needle holder, ultrasonic knife handle, ultrasonic knife transducer, etc.